Narishige Pc10 Manual New |verified| Jun 2026
The device uses gravitational force as its pulling mechanism, which can be adjusted through several manual and digital controls:
Decrease the Magnet dial value and slightly lower the Heater 2 temperature. This allows the glass to separate cleanly without stretching into a fine hair. 6. Maintenance and Cleaning Protocol
However, there is a persistent problem: the original paper manuals are often lost, damaged by chemical spills, or written in dense technical Japanese. This leads to a frantic search for a new Narishige PC-10 manual—not necessarily a brand-new machine, but a fresh, readable, and complete version of the documentation. narishige pc10 manual new
Operation centers on balancing heat and pulling force to create the desired pipette. This is controlled by two primary parameters:
| Problem | Probable Cause(s) | Remedy(ies) | | :--- | :--- | :--- | | No display on "HEATER LEVEL" when power is on. | - AC power cord is disconnected. - Fuses in the AC Power Inlet are blown. - Internal controlling board is malfunctioning. | - Firmly plug in the AC power cord. - Check and replace the fuse(s) with the correct rating. | | Pipette shape is different from usual, despite same settings. | - Heating element is deteriorated or contaminated. - Room temperature, humidity, or airflow (from AC) affecting the heating function. | - Increase the heater value to compensate. - Polish the mounting plates with sandpaper. - Move the instrument away from drafts. | | Pipette is distorted or off-centered. | - Glass capillary is not centered in the heating coil. - Heating coil coils are touching each other. | - Carefully re-center the capillary. - Use a wrench to separate the coils or replace the heater. | | Power blows fuse as soon as the device is turned on. | - A non-standard heating element is installed. - Heater values are set excessively high. | - Use only standard Narishige PC-10H elements. - Reduce heater value settings. | The device uses gravitational force as its pulling
Select the two-stage mode for precise control over the taper length and tip diameter.
For years, the PC-10 was the go-to tool for electrophysiologists. It was designed to replace older models like the and PB-7 , offering a more automated, dual-stage system. Maintenance and Cleaning Protocol However, there is a
This guide provides a detailed overview of the PC-10's features, setup, operation, and optimization techniques based on the standard operating procedures. 1. Introduction to the Narishige PC-10
These control the pulling force. For a standard sharp tip, set the magnets to a higher value for the initial pull and a lower value for the final pull. Step 4: Activating the Pull Unlock the safety stopper. Press the start button .
: Users have reported variability in tip size and resistance, requiring frequent manual adjustments to maintain consistency.
Select the desired number of . Adding weights creates shorter, stiffer tapers with larger tip diameters; removing weights yields longer, flexible tapers with smaller tip openings. Step 4: Initiating the Pull Turn on the main power switch.